Test Code LAB188 Lactate Dehydrogenase, Body Fluid
Performing Laboratory
Barnes-Jewish Hospital Laboratory
Methodology
Enzymatic (Lactate to Pyruvate)
Literature References:
Package insert: Lactate dehydrogenase LDH IFCC reagent. Roche Corporation, 03/2022.
Specimen Requirements
Specimen Type: Pleural fluid, Peritoneal fluid, Pericardial fluid
Container/Tube: Mint Green top tube
Specimen Volume: 2 mL
Additional Information: Indicate fluid type on specimen container.
Day(s) Test Set Up
Monday through Sunday
Turnaround Time:
STAT: 1 hour after receipt in laboratory
Routine: 2 to 4 hours after receipt in laboratory
Test Classification and CPT Coding
83615
Reference Values
Ratio of pleural to serum LD ≥ 0.6 is indicative of exudate and <0.6 of transudates.
Ratio of peritoneal to serum LD ≥ 0.6 is indicative of exudate and <0.6 of transudates.
Ratio of pericardial to serum LD ≥ 0.6 is indicative of exudates and transudates <0.6.
Literature References:
Balfe, Alan et al. The Biochemistry of Body Fluids. Association of Clinical Biochemists in Ireland. 1-35. October 2009.
Block, Darci R. Tietz Textbook of clinical Chemistry and Molecular Diagnositics. Elsevier Sixth Edition 2017. Body Fluids Abstract p. 925
Yataco, Jose C, Cleveland Clinic Journal of Medicine. Pleural effusions: Evaluation and Management. October 2005. Volume 72, Number 10.
Specimen Transport Temperature
Ambient
Additional Information
For BJH Laboratory Use Only
Alternate Tube Type: Red top tube
Analyte Stability:
See Analyte Stability Chart under General Information for a complete list.
Laboratory Processing Instructions:
Test performed in BJH Core lab. If specimen is received in a Vacutainer, check procedure to confirm suitability. Centrifuge upon receipt. Apply blue tape to the specimen aliquot tube. Deliver to the Cobas 8000. Completed specimen is stored frozen for 30 days.
LOINC Code Information
Fluid Type |
Loinc code |
Pleural Fluid |
60022-1 |
Peritoneal Fluid |
68387-0 |
Pericardial Fluid |
60019-7 |