Sign in →

Test Code LAB8146 CYP2C19

Performing Laboratory

Barnes Jewish Hospital Laboratory

Methodology

The detection of the *2, *3, and *17 alleles of CYP2C19 is determined by using polymerase chain reaction (PCR) and fluorescent probe detection automated by the Genomadix Cube CYP2C19 system. 

Note: This test identifies and reports only the CYP2C19 *2, *3, and *17 alleles. Other CYP2C19 alleles, which are rare, or unknown, will not be identified or reported. These other alleles may result in either a *1 call (wild-type), a no-call, or a call of a genetically related allele included in this system.

Specimen Requirements

Forms: Molecular Diagnostics Requisition

 

Specimen Type: Whole blood

Container/Tube: EDTA (lavender or pink) vacutainer tube

Specimen Volume: 2mL

Specimen Minimum Volume: 1 mL

Collection Instructions: Invert several times to mix blood. Clotted blood is not acceptable.

Day(s) Test Set Up

Monday through Sunday

Test Classification and CPT Coding

81225

Reference Values

Metabolizer Phenotype

CYP2C19 Enzyme Activity

Genotypes

Normal Metabolizer (NM)

Normal

*1/*1

Intermediate Metabolizer (IM)

Reduced

*1/*2, *1/*3, *2/*17, *3/*17

Poor Metabolizer (PM)

Very Low or Absent

*2/*2, *2/*3, *3/*3

Rapid Metabolizer (RM)

Increased

*1/*17

Ultra-Rapid Metabolizer (UM)

Very Increased

*17/*17

 

Specimen Transport Temperature

Samples must be sent at ambient temperature within 4 hours or refrigerated for up to 168 hours.

Additional Information

For BJH Laboratory Use Only
Analyte Stability:

Blood is stable up to 4 hours at ambient temperature and up to 168 hours refrigerated.