Test Code LAB5796 Epstein-Barr Virus (EBV) DNA Detection (Quantitative), plasma
Methodology
Testing is performed by real-time PCR on the cobas® 6800 platform using the cobas® EBV assay (Roche Molecular Systems, Inc.).
Specimen Requirements
Specimen Type: Plasma.
Container/Tube: Lavender or pink top (EDTA).
Specimen Minimum Volume: Minimum of 2 mL of whole blood in a dedicated lavender or pink top (EDTA) tube.
**Facilities outside of BJH should submit a minimum of 1 mL of plasma.
Collection Instructions:
1. Spin down and separate plasma from cells within 24 hours of collection.
2. Plasma should be stored at 2-8°C or -20°C upon separation. Please see Specimen Stability Information for additional information.
Day(s) Test Set Up
Monday, Wednesday, Friday
Turnaround Time (STAT, Routine)
STAT: Director approval required.
Routine: 24 – 48 hours
Test Classification and CPT Code
Test Classification:
This assay is FDA cleared for quantitative EBV DNA testing on plasma and its performance has been evaluated by the Barnes-Jewish Hospital Molecular Infectious Disease Laboratory.
CPT: 87799
Reference Values
Not detected.
Specimen Transport Instructions
Whole blood specimens should be delivered to the BJH Molecular Infectious Disease laboratory within 4 hours of collection for processing.
Specimen Stability Information
Specimens may be stored under the following conditions:
Prior to separation:
- 2 to 25°C for up to 24 hours
Following separation of plasma from whole blood:
- Room temperature for up to 24 hours
- 2 to 8°C for up to 6 days
- -20°C to -80°C for up to 6 months
Additional Information
Assay Limit of Detection - Plasma: 18.8 IU/mL
Assay Quantitative Range - Plasma: 35 IU/mL - 1.0x108 IU/mL
For BJH Laboratory Use Only
Laboratory Processing Instructions
Specimens should be delivered to the BJH MID laboratory. Specimen is stored frozen at the BJH MID lab for 21 days.