Test Code LAB919 HIV-1 RNA Detection and Quantification
Performing Laboratory
BJH Molecular Infectious Disease Laboratory
Methodology
Testing is performed by real-time PCR on the cobas® 6800 platform using the cobas® HIV-1 Assay (Roche Molecular Systems, Inc.).
Specimen Requirements
Specimen Type: Plasma
Collection Container/Tube: Lavender or pink top (EDTA)
Specimen Minimum Volume: Minimum of 2 mL of whole blood in a dedicated lavender or pink top (EDTA) tube.
**Facilities outside of BJH should submit a minimum of 1 mL of plasma.
Collection Instructions:
- Spin down and separate plasma from cells within 24 hours of collection.
- Plasma should be stored at 2-8°C or -20°C upon separation. Please see Specimen Stability Information for additional details.
Day(s) Test Set Up
Monday through Friday
Turnaround Time:
STAT: Director approval required
Routine: 2 to 3 business days
Reference Values
Not Detected.
Additional Information
Assay Limit of Detection
20 copies/mL
Assay Quantitative Range
20 copies/mL - 10,000,000 copies/mL.
For BJH Laboratory Use Only
Laboratory Processing Instructions:
Specimens should be delivered to the BJH MID laboratory. Specimen is stored frozen at the BJH MID lab for 21 days.
Specimen Stability Information
Plasma specimens may be stored under the following conditions:
Prior to separation:
- Room temperature for up to 24 hours
Following separation of plasma from whole blood:
- 2 to 8°C for up to 6 days
- -20°C to -80°C for up to 12 weeks
Test Classification and CPT Code
This assay is FDA cleared for quantitative HIV-1 RNA testing on plasma and its performance has been evaluated by the Barnes-Jewish Hospital Molecular Infectious Disease Laboratory.
CPT: 87536
Specimen Transport Instructions
Whole blood specimens should be delivered to the BJH MID laboratory within 24 hours of collection for processing.