Test Code Lab496 Rubella Antibodies, IgG, Serum

Infectious

Performing Laboratory

Barnes-Jewish Hospital Laboratory

Methodology

Multiplex Flourescent Immunoassay

Note: The use of this test to diagnose recent infection by testing acute and convalescent specimens is not recommended. The reported IgG level cannot be correlated to an endpoint titer.

Literature References:

Package Insert: MMRV IgG package insert, Bio-Rad Laboratories, Inc. 02/2019

Specimen Requirements

Specimen Type: Serum
Container/Tube:  Red top, Red/grey SST or Gold SST
Specimen Volume: Full tube, minimum 1mL of serum

 

Day(s) Test Set Up

Monday-Sunday 0700 - 1300

Turnaround Time:
STAT: not available
Routine: results available the same day of collection

Test Classification and CPT Coding

86762

Interpretive Data

Rubella IgG

Nonreactive:  No detectable antibody to rubella.  Such individuals are presumed to be uninfected with rubella and to be susceptible to primary infection.

Equivocal: Presence or absence of detectable antibody to rubella cannot be determined. Submit new specimen if clinically indicated.   

Reactive: Results suggest response to immunization or prior exposure to the virus.

 

Specimen Transport Temperature

Ambient

Additional Information

For BJH Laboratory Use Only

Analyte Stability:

Specimen may be stored at 2° C to 8° C for 7 days with no more than 3 freeze/thaw cycles.

Specimen may be frozen at -20° C for 6 months.

Laboratory Processing Instructions:

Test performed in BJH Core Lab.  Processed by automated line for delivery AOB with a minimum of 1 mL of plasma/serum. Completed specimens are stored refrigerated for 7 days.

LOINC Code Information

Name LOINC Code
Rubella Antibodies, IgG, Serum 22496-6